ICP-OES MICS全光谱直读电感耦合等离子体发射光谱仪

询价单
  • 介绍
  • 论点

P绩效指标

D检测器

高效的半导体冷却CCD固体探测器,在光谱仪的波长范围内具有连续像素,允许任意波长选择,并设计有自然防溢出功能

(1) D检测装置300多万个检测单元,读取速度≥2MHz

(2) 像素分辨率≤0.003nm

(3) 探测器制冷系统为了实现最低的探测器暗电流,使用了高效的三级半导体冷却

(4) 制冷温度-45℃达到工作温度的时间< 3分钟

(5) 光学系统恒温驱动步进光谱系统

(6) M单色仪Mid step grating and prism two-dimensional dispersion systemHigh energy,to ensure the stability of instrument testing, internal optical components such as gratings and prisms are fixed in position, and all elements are measured simultaneously in one exposure across the full wavelength range of the spectrometer

(7) 十、-ray roomxEquipped with precision light chamber constant temperature 35℃-40℃Argon or nitrogen can be used for purging the light chamberWhen measuring the<200nm spectral line, the gas       displacement is<3L/min

(8) Wavelength coverage165-900nmFull wavelength coverage

(9) ptical resolutionFHW):As189.042nm half width<0.007nmCa393.366nm half width<0.017nmBa614.172 half width<0.024nm, K766.490nm half width<0.035nm,To ensure the stability and optimal luminous flux of the optical system

Plasma

(1) Plasma observation methodVertical placement of torch tube

(2) RF generatorSolid-state generatorThe plasma coil is designed with a PTFE protective layer, which  is corrosion-resistant and maintenance free

(3) Frequency27.12MHZ

(4) RF power500-1600W

(5) Pneumatic controlConfigure 3 high-precision quality flow controllers, directly controlled by ICP-OES   software, including cooling gas, auxiliary gas, and atomization gasPrecision0.01 L/min

(6) Peristaltic pump5-channel 16-roller peristaltic pump

Performance analysis 

(1) A.nalysis speed≥50 elements or spectral lines per minuteAnd the integration time of each measured spectral line≥10s

(2) Sample consumption<2mlDetermine more than 70 elements

(3) Spectral line flexibilityQualitative, semi-quantitative and quantitative analysis can be conducted on any spectral line of the analyzed element, which is convenient for analysis and research

(4) Measure the linear dynamic range of spectral lines≥106Measured with Mn257.6nm, correlation     coefficient ≥0.9996),Provide supporting documents

(5) Internal standard calibrationSimultaneous internal standard calibration, which means that both internal standard elements and measurement elements must be exposed simultaneously

(6) PrecisionDetermination of 1ppm or 10ppm multi-element mixed standard solutionRSD ≤ 0.5% for   repeated measurements of ten times

StabilityDetermination of 1ppm or 10ppm multi-element mixed standard solutiondo not use internal standard calibrationContinuous 4-hour long-term stability RSD<1.0%